Up Product - Next Product
Patient Monitoring System (BW3F)
 

Patient Monitoring System (BW3F)

 
Details:
 
◆ Attractive streamline outline design, portable, compact and lightweight
◆ 7.0-inch color TFT LCD display
◆ Operation menu with multi-language interface selection: English, Spanish, Portuguese,  Chinese, French, German ……
◆ Big Font Display
◆ ST segment Analysis and Apnea Alarm function
◆ Drug Dose Calculation and OXCRG Software
◆ 72-hour tabular and graphic trends and data storage
◆ High energy Li-ion battery for short-term transport and temporary power-down
◆ Networking Capability. WAN communication function to network with central monitoring system and make  long-distance monitoring, diagnosis, maintenance and software upgrade possible
◆ Linux Operation System
◆ External display(VGA interface)
◆ Dynamic Monitoring of blood pressure
◆ Intelligent audio and visual alarm
◆ Application: Adult, Pediatric and Neonate
◆ Pacemaker detection and Arrhythmia Analysis( Optional)
 
Technical Specs
 
ECG
Input: 5 lead ECG cable 
Lead section: lead Ⅰ,Ⅱ,Ⅲ/ aVR, aVL, aVF/ V
Gain(mm/mV): 1/4,1/2,1,2,4
Frequency response:     Diagnosis          0.05 - 100Hz
                     Monitor mode       0.5 - 40 Hz
                     Surgery            1.0 - 25 Hz
Sweep speed (mm/sec): 12.5, 25, 50
Heart rate range: 20-254 BPM
ST segment deviation analysis: -0.8-+0.8mV
Protection: Against electrosurgical interference and defibrillation
 
NIBP
Method:          Automatic Oscillometric
Measurement range: Systolic  4.0 - 37.0 kPa
                 Diastolic 1.3 - 33.0 kPa
                 Mean 2.6 - 35.0 kPa
Scale: mmHg/Kpa
Operating Mode: Manual/Auto/STAT
Over pressure Protection:  Yes
 
SPO2
Measurement range: 0 - 100 %
Pulse Rate range:   30- 254BPM
Accuracy:         ± 2 %
Probe:            For Neonate, pediatric, and adult
 
Temperature
Channel: Two Channels
Measurement range: 29- 45℃
Accuracy: ±0.1℃
Scale: Celsius/Fahrenheit
 
RESP.
Method: Impedance
Measurement range: 0 - 120 BPM
Lead Connection: RA & LL or RA & RA
 
Power requirements 
Input: 98~260 V AC, 50/60Hz
Consumption: ≤  80VA 
 
Safety standards:  GB9706.1  IEC  60601-1 UL  544
 
EtCO2( Optional)
Method: Side Stream
Measurement range: 0 % - 10.0 % in CO2 and et CO2)
Display: Waveform, Respiration Rate, EtCO2, FiCO2
Scale: mmHg/Kpa
 
2-IBP ( Optional)
Channel: 2 Channels
Parameter: SYS/DIA/MAP/Pulse Rate
Scale: mmHg/kPa
Related Products:
ClosePrint